EU-2000 (ANAPOL)

CO
Electrochemical
Range : 0 – 2'000 ppm
Resolution : 1 ppm
Response time : ∆ 90% < = 45 seconds
Error limit : ± 0.10 x display or ± 12 ppm

NO (option)
Electrochemical
Range : 0 - 2'000 ppm
Resolution : 1 ppm
Response time : ∆ 90% < = 45 seconds
Error limit : ± 0.1 x display or ± 10 ppm

SO2 (option)
Electrochemical
Range : 0 - 2'000 ppm
Resolution : 1 ppm
Response time : ∆ 90% < = 45 seconds

TG (gas temperature)
NiCr/Ni thermocouple
Range : 0 - 450°C (800°C Option)
Error limit :

Temperature Device Sensor Total
0 - 100 °C ± 1 °C ± 2 °C ± 3 °C checked by metas
100 - 200 °C ± 1 % ± 2 % ± 3 % checked by metas
200 - 350 °C ± 2 °C ± 4 °C ± 6 °C checked by metas

TA (fresh air temperature)
NiCr/Ni thermocouple
Range : 0 - 80°C
Error limit : ± 3 °C max

Draft (draft metering)
Semi-conductor DMS bridge
Range : - 4 / + 50 hPa

Calculations
CO2 0 - 20,0% vol
qA 0 - 99,9%
rend 99,9%-0
la (Lambda) 1 - ∞
NOx NO + NO2 = NOx

Storage temperature :
-20 to 50 °C

Display
LCD/Dot Matrix
4 lines of 20 characters

Printer
Dot Matrix Printer
24 characters per line

Power supply
Mains voltage : 85 – 264 VAC / 47-63 Hz
Batteries LiIon 12V DC – 4 Ah

Calibration
After successful leak test, 100 sec.

Soot metering
Filter paper method
1.63 ± 0.11 l (6 mm probe)

Hose connection
3.5 m

Dimensions / weight
460 x 370 x 150 mm
6.6 kg

 
Equipment has always been an important part, having equipment that is technologically advanced gas analyzer can make the job easier. In terms of precision gas analysis, and the accuracy of the data becomes very important, therefore we as a supplier of gas analyzer Indonesia equipment very understanding so we always give priority to products that have the precision and accuracy of the data from the gas analyzer.